DatasheetQ Logo
Electronic component search and free download site. Transistors,MosFET ,Diode,Integrated circuits

AD8220ARMZ-RL View Datasheet(PDF) - Analog Devices

Part Name
Description
Manufacturer
AD8220ARMZ-RL Datasheet PDF : 28 Pages
First Prev 21 22 23 24 25 26 27 28
ELECTROCARDIOGRAM SIGNAL CONDITIONING
The AD8220 makes an excellent input amplifier for next
generation ECGs. Its small size, high CMRR over frequency,
rail-to-rail output, and JFET inputs are well suited for this
application. Potentials measured on the skin range from 0.2 mV
to 2 mV. The AD8220 solves many of the typical challenges of
measuring these body surface potentials. The AD8220’s high
CMRR helps reject common-mode signals that come in the
form of line noise or high frequency EMI from equipment in
the operating room. Its rail-to-rail output offers wide dynamic
range allowing for higher gains than would be possible using
other instrumentation amplifiers. JFET inputs offer a large
input capacitance of 5 pF. A natural RC filter is formed reducing
high frequency noise when series input resistors are used in
front of the AD8220 (see the RF Interference section).
AD8220
In addition, the AD8220 JFET inputs have ultralow input bias
current and no current noise, making it useful for ECG
applications where there are often large impedances. The MSOP
package and the AD8220’s optimal pinout allow smaller
footprints and more efficient layout, paving the way for next
generation portable ECGs.
Figure 67 shows an example ECG schematic. Following the
AD8220 is a 0.03 Hz high-pass filter, formed by the 4.7 μF
capacitor and the 1 MΩ resistor, which removes the dc offset
that develops between the electrodes. An additional gain of 50,
provided by the AD8618, makes use of the 0 V to 5 V input
range of the ADC. An active, fifth order, low-pass Bessel filter
removes signals greater than approximately 160 Hz. An OP2177
buffers, inverts, and gains the common-mode voltage taken at
the mid-point of the AD8220 gain setting resistors. This right
leg drive circuit helps cancel common-mode signals by
inverting the common-mode signal and driving it back into the
body. A 499 kΩ series resistor at the output of the OP2177
limits the current driven into the body.
AB
C
15k
2.2pF
10k
10pF
10k
2.2pF
INSTRUMENTATION
AMPLIFIER
+5V
G = +14
1.18k
2.5V
AD8220 HIGH-PASS FILTER 0.033Hz
+5V
24.9k
–5V +5V 24.9k
4.12k
4.7µF
G = +50
57.6k14k
+5V
AD8618
–5V
+5V
220pF
1M
LOW-PASS FIFTH ORDER FILTER AT 157Hz
14k19.3k
33nF
14.5k
68nF
47nF
+5V
AD8618
+5V
19.3k
AD8618
1.15k
AD8618
+5V
14.5k
500
2.7nF
33nF
4.99k
22nF
–5V
OP2177
2.5V
2.5V
2.5V
2.5V
–5V
OP AMPS
AD7685
ADC
REF
+5V
4.7µF
REFERENCE
ADR435
68pF
12.7k
866k
+5V
499k
OP2177
–5V
Figure 67. Example ECG Schematic